Section 9 Details on Device Messages
9-32
9.22 DCB [Data Condition Memory B]
Function
Reads data measurement conditions from memory B.
Header Program Query Response
DCB None DCB ? λ 1, λ 2, n
Value of λ 1
λ 1 indicates a start wavelength. The unit is always nm. A value is output down
to the second decimal place.
Value of λ 2
λ 2 indicates a start wavelength. The unit is always nm. A value is output down
to the second decimal place.
Value of n
n indicates a sampling point, 251, 501, 1001, 2001, or 5001.
9.23 DEL [FD File Delete]
Function
Deletes the specified file from the FD.
When file deletion is complete, bit 2 (transfer end bit) of the extended event status
register (ESR2) is set to 1.
Header Program Query Response
DEL DEL n None None
Value of n
Input a DOS-recognizable file name comprising eight or fewer characters in the
following format:
xxxxxxxx.dat
The file extension “.dat” may be omitted.