9-57
9.71 SAV [FD File Save]
■ Function
Saves measurement data in the specified file on the FD.
When file saving is complete, bit 2 (transfer end bit) of the extended event status
register (ESR2) is set to 1.
Header Program Query Response
SAV SAV n None None
■ • Value of n
Input a DOS-recognizable file name comprising eight or fewer characters in the
following format:
xxxxxxxx.dat
The file extension “.dat” may be omitted.
9.72 SMT [Smooth]
■ Function
Sets a number of smoothing points.
Header Program Query Response
SMT SMT n SMT ? n
■ Value of n
The unit is number of points. Input one of the following values or OFF:
3, 5, 7, 9, 11, or OFF
■ Initial setting value
n is a backed up value.
■ Default
n = OFF
9.73 SPC [Spectrum]
■ Function
Goes to the spectrum measurement mode.
Header Program Query Response
SPC SPC None None
9.73 SPC [Spectrum]