Test Records A-5 MS2036C VNA Master Test Record
MS20xxC MM PN: 10580-00307 Rev. D A-11
Instrument Information
Spectrum Analyzer Amplitude Accuracy – MS2036C (continued)
Serial Number: Firmware Revision: Operator:
Options: Date:
Table A-20. Spectrum Analyzer Amplitude Accuracy Across Frequency Setup Table
Frequency
Required Sensor B reading for
–2 dBm @Attenuator output
Required Sensor B reading for
–30 dBm @Attenuator output
10.1 MHz dBm dBm
100 MHz dBm dBm
500 MHz dBm dBm
1000 MHz dBm dBm
2000 MHz dBm dBm
3000 MHz dBm dBm
4000 MHz dBm dBm
5000 MHz dBm dBm
6000 MHz dBm dBm
7000 MHz dBm dBm
8000 MHz dBm dBm
8999 MHz dBm dBm
Table A-21. Spectrum Analyzer Amplitude Accuracy Across Frequency – 10.1 MHz and 100 MHz
Freq
(MHz)
Test
Power
(dBm)
Atten
Level
(dB)
Spec
(dB)
Marker
M1
Reading
(dBm)
Dev
(dB)
Freq
(MHz)
Test
Power
(dBm)
Atten
Level
(dB)
Spec
(dB)
Marker
M1
Reading
(dBm)
Dev
(dB)
10.1
–30 0 ±1.3
100
–30 0 ±1.3
–305±1.3 –305±1.3
–30 10 ±1.3 –30 10 ±1.3
–30 20 ±1.3 –30 20 ±1.3
–2 30 ±1.3 –2 30 ±1.3
–2 40 ±1.3 –2 40 ±1.3
–2 50 ±1.3 –2 50 ±1.3
–2 60 ±1.3 –2 60 ±1.3