2-3 Transmission Dynamic Range Vector Network Analyzer Performance Verification
2-4 PN: 10580-00307 Rev. D MS20xxC MM
8. Follow the instructions of the software to connect the Open, Short, and Load of the calibration tees to
Port 1 and to the open end of the RF cable (extended Port 2) for the measurement calibration. Figure 2-1
shows the general calibration setup.
9. Follow the instructions of the software to perform measurements of the Transmission Noise Floor.
10. Pass/Fail status of the test is displayed on the computer. The software can also provide a hard copy
(printout) of the test result, if desired.
11. Attach the printout of the test results to the test record in Appendix A:
MS2026C – Section “VNA Transmission Dynamic Range – MS2026C” on page A-2
MS2027C – Section “VNA Transmission Dynamic Range – MS2027C” on page A-3
1. Test Port 1
2. Test Port 2
3. RF Test Cable
4. Open, Short, and Load of female OSL Calibration Tee
5. Open, Short, and Load of male OSL Calibration Tee
Figure 2-1. Calibration Setup
OPEN
SHORT
LOAD
OPEN
SHORT
LOAD
1
2
3
4
5
Port 1 Port 2