3-10 Third Order Intercept (TOI) Spectrum Analyzer Performance Verification
3-32 PN: 10580-00307 Rev. D MS20xxC MM
13. Turn RF to ON on MG3692X #1.
14. On the MS203xC, press the Shift key and then press the Trace (5) key.
15. Press the Trace A Operations soft key and then the # of Averages soft key.
16. Enter 2 and then press the Enter key to set # of Averages to 2.
17. Wait until the Trace Count displays 2/2.
18. Press the Marker hard key and then the Peak Search soft key.
19. Record the Marker amplitude reading at 2400.151 MHz in Table A-27, “Spectrum Analyzer Third Order
Intercept” on page A-14 for MS2036C, in Table A-51, “Spectrum Analyzer Third Order Intercept”
Figure 3-6. Third Order Intercept Measurement Setup
Signal Generator
Power Meter
10 MHz Reference
Power
Splitter
Attenuators
Signal Generator
Attenuators
6 dB
20 dB2 dB 20 dB 6 dB 2 dB
To
Rear
VNA Master
Adapter
34NN50A
Sensor

Table of Contents

Other manuals for Anritsu VNA Master MS2026C