EasyManua.ls Logo

Architect i2000SR - Page 178

Architect i2000SR
212 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
i2000SR interface module Service Manual
Page 178 of 212 625798100.APS.5.doc
STEPS REFERENCES
5 At the i2000SR SCC, perform Sample
and STAT probe calibrations
Maintenance, As Needed, the
appropriate probe to calibrate.
1111 Sample Probe Calibration
1117 STAT Probe Calibration
NOTE:
At the following screen prompt;
Use the (p/n 204021) LAS pipettor
calibration tool illustrated in the
reference area.
6 Place the LAS Probe calibration tool
(P/N 204021) into the IM track at the
appropriate sampling location.
Ensure that the tool [2] is pressed
firmly against the gate [1]. If it is not,
the alignment will be compromised.
7 Follow the i2000SR screen menu
options to select the LAS probe
alignment option.
8 Follow the screen menu options to
position the probe over the center of
the tool [2].
As the probe must be precisely
centered, a combination of probe and
gate position adjustments may be
required.

Table of Contents