Room Devices
Telligence
64PM000224A / 10 August 2022 / Ver. H
Figure 66: Cardiac and Emergency Pull Switches
A Cardiac and Emergency pull switches have no audio capability and cannot be used for voice
communication unless it is paired with a nearby speech module.
To activate or cancel a cardiac/emergency call event:
1. Pull the Cardiac or Emergency switch to send a Code Blue (cardiac) or emergency call event
to staff consoles and annunciators. Pulling the switches causes the following to occur:
Lights linked buttons illuminate to confirm call placement.
The call assurance tone sounds at any associated module or Room Display.
The event will display on the associated corridor and zone lights.
The pull tones sound at staff consoles and annunciators and at stations where staff
follow is active.
A call event appears on the Call screen on staff consoles, emergency annunciators and
unit annunciators.
2. To cancel the call, push the pull switch.
Code button lights extinguish.
Call lights extinguish on any associated Bedside module or staff device, if they are
linked.
Dome lights extinguish.
Tones stop at staff consoles, annunciators, and at stations where staff follow is active.
The nurse call event is removed from staff consoles and annunciators.
The 37-Pin Bed Receptacle
The 37-Pin Bed Receptacle is a wall-mounted, 37-pin D-sub female connector attached to a
stainless steel plate. The receptacle supports electronic beds with side-rail communications.