Cleaning and Maintenance
Telligence
70PM000224A / 10 August 2022 / Ver. H
4. Gently dry the surface of the touch screen using a dry, soft cloth of the same type (a lint-free
eyeglass-cleaning cloth or a microfiber cloth).
5. Press Exit Settings when finished.
Cleaning the Housing
WARNING: Never submerge a staff console, annunciator, TelliConnect Station, or any attached
cords.
To clean the housing:
1. Slightly moisten a lint-free eyeglass-cleaning cloth or soft microfiber cloth with an approved
cleaning solution.
CaviWipes (13-1100)
Cidex
Clorox Disinfecting Wipes (bleach free)
Clorox Germicidal Wipes (446035309)
Dispatch Disinfectant Wipes
Ecolab Achtichlor Plus
Ecolab Asepti-Wipe II (Product # 61027330)
Medtrol Gluco-Chlor Bleach (GC110)
PDI Sani-Cloth AF (Product #P84372) (white canister)
PDI Sani-Cloth AF3 (P13872) (gray canister)
PDI Sani-Cloth Bleach (H24795) & P54072 (orange canister)
PDI Sani-Cloth HB (Product # Q08472) (green canister)
PDI Sani-Cloth Plus (Product # Q89072) (red canister)
Staphine
Virkon S
10% bleach/water solution
2. Use the cloth to gently clean the surface of the housing being careful not to make contact
with the touch screen.
3. Use the moist cloth to gently clean the handset, handset cord, headset, and headset cord, if
these are present.
4. Use another soft, dry cloth to dry the surface of the housing.
Cleaning Patient Stations and Room Modules
Cleaning mode temporarily deactivates patient station and any attached room modules so that
they can be cleaned without accidentally changing the status of nurse calls during the cleaning
process.
Note: Since patient stations cannot be used to place nurse calls while they are in Cleaning
mode, it is best to clean them when no patients are occupying the beds in the room.
To clean patient stations and room modules:
1. Ensure that all calls on associated with the devices have been canceled.
2. Slightly moisten a lint-free eyeglass-cleaning cloth or soft microfiber cloth with an approved
cleaning solution.
Cavi Wipes
Cidex
Clorox Disinfectant wipes