SYSTEM CONFIGURATION Chapter 8
8 - 7
CAUTION
To prevent corruption of the configuration data, do not turn off the
pump while UPDATING is being displayed.
8.4.1.2.4 MODIFY LIBRARY
The MODIFY LIBRARY function allows the operator to rename, enable or disable a
protocol library.
Immediately after the MODIFY LIBRARY function is selected, the pump displays ENABLE
NONE LIBRARY OPTION with a Y or N response, indicating respectively that the option is
currently enabled or disabled. Enabling this option adds a NONE selection to the list of
available libraries, allowing protocol selection to be bypassed during normal programming.
Press or if necessary to change the response and press CONFIRM to complete this
step.
At the SELECT LIBRARY prompt, press or as necessary to select the library to
be modified, and press CONFIRM to complete the selection.
Once a library has been selected, the pump displays NAME <library-name> where <library-
name> is the name of the selected library. Press CONFIRM to accept the current library
name, or change the name to any name that is not already assigned to another library.
Press CONFIRM to complete the library name entry.
Next, the pump displays ENABLE NONE PROTOCOL OPTION with a Y or N response,
indicating respectively that the option is currently enabled or disabled. Enabling this option
adds a NONE selection to the list of available protocols in the library, allowing protocol
selection to be bypassed during normal programming. Press or if necessary to
change the response and press CONFIRM to complete this step.
Next, the pump displays ENABLE <library name> with a Y or N response, indicating
respectively that library is currently enabled or disabled. Only the names of enabled
libraries are displayed at the SELECT LIBRARY programming step. Press or as
necessary to change the response and press CONFIRM to complete this step. The pump
displays UPDATING.
CAUTION
To prevent corruption of the configuration data, do not turn off the
pump while UPDATING is being displayed.

Table of Contents

Other manuals for baxter Auto Syringe AS50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the baxter Auto Syringe AS50 and is the answer not in the manual?

baxter Auto Syringe AS50 Specifications

General IconGeneral
Brandbaxter
ModelAuto Syringe AS50
CategoryWater Pump
LanguageEnglish

Related product manuals