Accessories
104 Hillrom Progressa ICU Hospital Bed Instructions for Use (171528 REV 15)
To Remove
8. Make sure the power cord is connected to the Experience
Pod Device.
NOTE:
Do not power the Experience Pod Device by the auxiliary outlet.
9. Plug the Experience Pod Device power cord into AC power.
10. If applicable, plug in the bed power cord.
11. Lower the bed to its lowest position.
1. Make sure the bed’s brake is set, and a patient is not in the
bed.
2. Raise the bed to access underneath the head section.
3. Unplug the Experience Pod Device power cord from AC power.
4. Put the Experience Pod Device in the transport position as
shown.
5. From underneath the head end of the bed frame, remove the
retainer from the pin.
6. Remove the pin, and then remove the Experience Pod
Device from the bed.