pro+ Surface Instruction for Use (209196 REV 8) 101
Specifications
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration -
Electromagnetic Immunity
The pro+ surface is intended for use in the electromagnetic environment specified below.
The customer or the user should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment—
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 2 kV, 4 kV, ±8 kV,
and
± 15kV Air
± 8 kV Contact
± 2 kV, ± 4 kV, ±8
kV,
± 15 kV Air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV (100 kHz
repetition
frequency) for
Power Supply
Lines
± 2 kV (100 kHz
repetition
frequency) for
Power Supply
Lines
Mains power quality
should be that of a
typical hospital
environment.
Surge
IEC 61000-4-5
±0.5 kV, ±1 kV
Line(s) to Line(s)
± 0.5 kV, ±1 kV, ±2
kV Line(s) to
Ground
± 0.5 kV, ±1 kV
Line(s) to Line(s)
±0.5 kV, ±1 kV, ± 2
kV Line(s) to
Ground
Mains power quality
should be that of a
typical hospital
environment.
Voltage Dips, IEC
61000-4-11
< 0% U
T:
0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°,
and 315°
0% U
T
: 1cycle and
70%
U
T
: 25/30 cycles
Single phase: at 0°
(see note)
< 0% U
T
: 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°,
and 315°
0% U
T
: 1cycle and
70%
U
T
: 30 cycles
Single phase: at 0°
(see note)
Mains power quality
should be that of a
typical hospital
environment. If the user
of the pro+ surface
requires continued
operation during power
mains interruption, it is
recommended that the
pro+ surface be
powered from an
uninterpretable power
supply.
Voltage
interruption
IEC 61000-4-11
0% U
T
: 250/300
cycles
0% U
T
: 300 cycles
Power Frequency
Magnetic Field
IEC 61000-4-8
30 A/m
50 Hz and 60 Hz
30 A/m
50 Hz and 60 Hz
Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
hospital environment.
Proximity
Magnetic Fields IEC
61000-4-39
30 kHz, CW, 8 A/m
134.2 kHz, 50%
Pulse at 2.1 kHz, 65
A/m
13.56MHz, 50%
Pulse at 50 kHz, 7.5
A/m
30 kHz, CW, 8 A/m
134.2 kHz, 50%
Pulse at 2.1 kHz, 65
A/m
13.56MHz, 50%
Pulse at 50 kHz, 7.5
A/m