Chapter 4: Maintenance 107
NOTICE If you wish to run BD FACSRinse solution as part of the daily cleaning
procedure, repeat the daily cleaning using BD FACSRinse solution in wells A1–
A4 and DI water in wells B1–B4.
5 Remove or open the safety cover and place the plate on the plate holder.
Make sure the plate corresponds to the type selected in the software. Orient
the plate with well A1 on the back-right corner of the stage. See Figure 4-4
on page 107. Verify that the sample coupler is properly installed and not
leaking.
Figure 4-4 Orienting the plate
6 Replace the safety cover.
7 [BD LSR II] Place the cytometer in Run mode.
8 Verify that the Cytometer window displays Cytometer Connected.
To ensure that the 10% bleach solution retains its full germicidal effect,
prepare a fresh solution daily.