48 BD High Throughput Sampler User’s Guide
• Click on a well and then click Record Data under Basic Controls to start
recording data
NOTICE While running plate-based acquisition, the Start Acquisition on pointer
change preference in the User Preferences window is ignored and the software
proceeds as if the preference were set to Off.
NOTICE During manual acquisition, high throughput mode is not applicable.
Stopping time in manual acquisition mode is calculated the same as in automatic
sequence mode. The software will stop acquisition or recording of a well when
any of the following stopping rules have been met:
• the specified number of events were collected
• the stopping time was reached
• the file exceeded memory specifications
• acquisition was stopped by clicking the Stop Acquiring button under Basic
Controls
Once acquisition of the well is complete, any remaining sample is flushed from
the system. The controls in the Acquisition Dashboard are briefly disabled while
the loader is reset.
Analysis View Components
The Analysis view provides an interface for
• displaying keyword results in the plate layout
• reviewing and analyzing data collected from a plate
• starting a batch analysis
For information on performing these functions, see Analysis View Functions on
page 51.