162 BD High Throughput Sampler User’s Guide
Figure B-2 Drip reservoir containing absorbent pad [BD LSR II and BD FACSCanto only]
15 Place the sample tubing and attached sample coupler inside the drip
reservoir.
Cleaning the HTS Unit
You must decontaminate both the internal fluidics and external surfaces of your
HTS unit before it is repaired. If you were able to run a monthly clean before you
removed the HTS unit, the internal fluidics are decontaminated; skip to
Decontaminating External Surfaces on page 167. If you were not able to run a
monthly clean, proceed with the following section.
All cytometer surfaces that come in contact with biological specimens can
transmit potentially fatal disease. Use universal precautions when cleaning
the cytometer. Wear suitable protective clothing, eyewear, and gloves.
absorbent pad