116 BD High Throughput Sampler User’s Guide
Replacing the Sample Injection Tubing
The sample injection tubing connects the injection port/wash station to the
secondary pump valve tubing. Replace the tubing every 6 months.
Figure 4-5 Disconnecting sample injection tubing from injection port/wash station
1 Perform the monthly cleaning procedure to decontaminate the sample
injection port/wash station.
See Monthly Cleaning on page 111.
2 Turn off the HTS power switch (BD LSR II and BD FACSCanto), or turn
off the cytometer (BD FACSCanto II).
3 For easy access, remove the HTS unit.
See Removing the HTS Unit on page 159 for instructions.
4 Move the probe arm up, and then push it toward the front of the HTS.
All cytometer surfaces that come in contact with biological specimens can
transmit potentially fatal disease. Use universal precautions when changing
tubing. Wear suitable protective clothing and gloves.