122 BD High Throughput Sampler User’s Guide
2 Place the HTS unit on its side.
NOTICE To avoid scratching the finish, handle the HTS with care. Do not
drag the unit on the laboratory bench.
3 Remove the filter retainer.
The filter retainer is located within the air filter element in the base plate of
the unit.
• Place your middle and index fingers of each hand on the opposite edges
of the filter retainer; place your thumbs on the fan guard through the
retainer apertures.
• Support the fan guard with your thumbs as you pull out the retainer
with your fingers.
NOTICE To prevent the HTS unit from tipping, use your thumbs to hold
the unit in place while pulling.
filter
retainer
Example of filter for BD LSR II and BD FACSCanto