Chapter 4: Maintenance 141
4 Verify that the sample probe properly performs the following sequence:
• The probe travels to the home position.
• The probe moves up, and then moves to well A1.
The following message appears.
5 Remove or open the safety cover and verify that the probe is in the center
of the nearest well (position A1).
6 Replace or close the safety cover and click OK.
The probe moves to well H12 and the following message appears.
7 Remove or open the safety cover and verify that the probe is in the center
of the farthest well (position H12).
8 Replace or close the safety cover and click OK.
The probe returns to the injection port and the following message appears.
9 Click OK to dismiss the message.