168 BD High Throughput Sampler User’s Guide
Figure B-7 HTS unit, top view
3 Repeat step 2 moving the plate holder to the far right, front position, and
then back position.
4 Clean the remaining external surfaces of the cytometer using paper towels
saturated with a 10% bleach solution.
Make sure all surfaces are clean.
5 Dispose of the paper towels used for cleaning as biohazardous waste.
6 Rinse all surfaces by repeating steps 1 through 5, substituting DI water for
the bleach solution.
7 Allow the cytometer to dry thoroughly.
plate holder
stage