96 BD High Throughput Sampler User’s Guide
Figure 3-20 HTS unit detached from cytometer
9 If applicable, perform the monthly cleaning as outlined in Monthly
Cleaning on page 111, install a new sheath filter, and fill the sheath tank
with BD FACSFlow solution.
Exchanging the Sheath Fluid on a BD LSR II
Follow the instructions under BD LSR II Monthly Cleaning on page 111. In
step 14, replace the sheath filter with a new filter instead of reconnecting the old
filter.
NOTICE If you exchanged BD FACSFlow with BD FACS sheath solution with
surfactant, install the HTS sample coupler, then prime the system twice by
choosing HTS > Prime. This will exchange the fluid in the syringes with BD FACS
sheath solution with surfactant.
cytometer
thumb screw
thumb screw
support bracket
base plate