EasyManua.ls Logo

BERTHOLD TECHNOLOGIES LB4710-080 - Page 221

Default Icon
227 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 8 Appendix
Mini-Switch LB 471
76
Professional Mode Parameters
Code
no.
Designation
Value range
Factory
setting
Device
setting
Page
00
Password
0000 - 9999
21
01
Year
1970 - 2099
Current year
21
02
Month / Day
01.01 - 12.31
Current date
22
03
Hour / Minute
00.00 - 23.59
Current time
22
04
Operation mode Standard/Professional
0 - 1
22
05
Detector code
0 - 99
99 GMZ
0 FSK
22
06
Nuclide: 0=Co-60, 1=Cs-137
0 - 1
0
23
07
Automatic password protection
0 - 9999
0
23
08
Warning relay as second alarm relay
0 1
10 - 100
0
23
09
Alarm relay follows the error relay
0 - 1
0
25
10
Reading (%)
-999 - 9999
Reading
25
11
Reading averaged
0 - 999.9
Reading
25
12
Time constant (s)
0.1 999.9
-1
25
13
Live count rate
0 - 999.9
Reading
26
14
Maximum time constant (s)
0 - 999
999
26
15
Standard reading
10 - 11
10
26
16
Max. or min. limit value switch
0=Max, 1=Min
0 - 1
0
27
17
Switching threshold (%)
0 - 100
-1
27
18
Switching threshold in (cps)
0 - 999.9
-1
28
19
Hysteresis (%)
0 - 999
-1
29
20
Empty count rate (no input)
0 - 999.9
Reading
19
21
Full count rate (no input)
0 - 999.9
Reading
29
22
Zero count rate (no input)
0 - 9.999
Reading
30
30
Empty count rate
0 - 999.9
20 GMZ
300 FSK
31
31
Full count rate
0 - 999.9
-1
31
32
Zero count rate
0 - 9.999
Depending on
detector code
32
33
Measuring path (in mm)
0 - 9999
0
33
34
Gas density (kg/m³)
0 - 9999
0
33
35
Bulk density (kg/m³)
0 - 9999
0
34
36
Compute
35.01 35.08
Reading
35
37
Counting time for calibration (s)
5 - 599
60
36
38
Bulk cone diameter (mm)
0 - 9999
0
37
39
Half-value layers
1 - 9
2
36

Table of Contents

Related product manuals