Patient monitor User’s manual
13
Measurement time
Normally, it is 20s to 45s (depending on HR and moving
interference typically)
Maximal measurement time: 120s(adult / pediatric), 90s
(neonate)
Initial inflation
pressure
Adult default: 160 mmHg
Pediatric default: 130 mmHg
Neonatal default: 75 mmHg
Software overpressure
protection
Double hardware,software overpressure protection
Adult: (297±3) mmHg
Pediatric: (147±3) mmHg
Neonatal: (252±3) mmHg
Assistant venipuncture
inflation mode
Inflation pressure(should be close to diastolic pressure):
Adult : 20 mmHg to 120 mmHg (normally 80 mmHg);
Pediatric: 20 mmHg to 80 mmHg (normally 60mmHg);
Neonatal: 20 mmHg to 50 mmHg (normally 40mmHg).
Intervals for periodic
measurement time
1min, 2min, 3min, 4min, 10min, 15min, 30min, 45min,60min,
90min, 120min,240min, 480min.
Alarm range
Sys
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.
Dia
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.
Map
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.
Alarm Indication
Blinking display of the data and parameters, text prompts,
Three levels of alarming: sound-light alarming, alarming with
blinked data and parameters, and that with text prompts.
Work mode
Adult Single, Cycle, STAT
Pediatric Single, Cycle, STAT
Neonatal: Single, Cycle
PR
PR range 40 bpm to 240 bpm
Resolution 1 bpm
Recovery time after <5s