EasyManua.ls Logo

Biolight Q7 - Setting CO

Default Icon
210 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor User’s manual
13-6
13.4 Setting CO
2
13.4.1 Setting CO2 Parameter
Select the CO
2
parameter area to enter CO
2
parameter setting menu.
Setting Unit
Select【Unit】in the CO
2
parameter setting menu, and the options are
【mmHg】,【kPa】or【%】.
Setting Apnea Alarm Time
For mainstream CO
2
and microstream CO
2
module, select【Apnea Alm】in
the CO2 parameter setting menu, you can select the apnea time as required in
the options. The monitor indicates an alarm when a pre-adjusted time has
elapsed since the last detected breath.
Setting Expiring Cycle
The value of EtCO
2
and FiCO
2
in the CO
2
parameter area will be refreshed
in real-time. As for the mainstream CO
2
module and microstream CO
2
module,
you can set the method to calculate EtCO
2
and FiCO
2.
The【Et Cycle】can be set in the CO
2
parameter setting menu:
【One Breath】:calculate the EtCO
2
and FiCO
2
by every respiratory wave.
【 10s 】 or 【 20s 】: a time interval, during which the maximum CO
2
concentration is EtCO
2
, and the minimum CO
2
concentration is FiCO
2.
Setting Alarm Switch
Select 【Alarm Switch】in CO
2
parameter setting. Alarm ON or OFF can be
set.
Setting Alarm Source
Select the alarm source for CO
2
. When the selected parameter or one of the
parameter exceed alarm limit, the monitor will give out alarm signal. The
options are as follows:
【Et】:Only EtCO
2
exceeds the alarm limit will trigger the alarm.
【Fi】:Only FiCO
2
exceeds the alarm limit will trigger the alarm.
【awRR】:Only awRR exceeds the alarm limit will trigger the alarm.

Table of Contents