Patient Monitor User’s manual
2-3
2.2 General Safety
Warning: The monitor is neither a therapeutic instrument nor a
device that can be used at home.
1.Safety precautions for installation
Connect the power cord to a properly earthing socket. Avoid putting the
socket used for it in the same loop of such devices as the air conditioners,
which regularly switch between on and off.
Avoid putting the monitor in the locations where it easily shakes or
wobbles.
Enough space shall be left around the monitor so as to guarantee normal
ventilation.
Make sure the ambient temperature and humidity are stable and avoid the
occurrence of condensation in the operation process of the monitor.
Warning: Never install the monitor in an environment where
flammable anesthetic gas is present.
2. Monitor conforms to the safety requirements of IEC 60601-1. This monitor
is protected against defibrillation effects.
3. Notes on symbols related to safety
Type BF applied part, defibrillation protected
The unit displaying this symbol contains an F-Type isolated
(floating) applied part providing a high degree of protection
against shock, and is defibrillator-proof.
Type CF applied part, defibrillation protected
The unit displaying this symbol contains an F-Type isolated
(floating) applied part providing a high degree of protection
against shock, and is defibrillator-proof.
The type CF applied parts provide a higher degree of protection
against electric shock than that provided by type BF applied
parts.