Patient Monitor Users manual
11-4
Auto mode, the first auto measurement will be started within 5min, and then
the monitor will start the measurement continually repeated in the set interval
after the first measurement.
Warning: Prolonged NIBP measurements in Auto mode are associated
with purport, ischemia and neuropathy in the limb wearing
the cuff. When monitoring a patient, examine the
extremities of the limb frequently for normal color, warmth
and sensitivity. If any abnormality is observed, stop the
NIBP measurements.
11.5.4 STAT Measurement
Select smartkey, or select NIBP parameter area, and selectStart/Stop
STATto start or stop the STAT measurement. The measurement will last
5min.
11.6 NIBP Display
There is no waveform displayed for NIBP measurement, the NIBP
readings are displayed in the parameter area. The following figure shows the
NIBP display screen, the display on your monitor may be looked slightly
different.
1. Pressure unit
2. Measurement mode
3. Mean arterial blood pressure
4. Diastolic blood pressure
5. Systolic blood pressure
1
2
3
4
5