Patient Monitor Users manual
2-7
Loss of data
Should the monitor at any time temporarily lose patient data, close patient
observation or alternative monitoring devices should be used until monitor
function is restored.
If the monitor does not automatically resume operation within 60s, restart
the monitor using the power switch. Once monitoring is restored, you should
verify correct monitoring state and alarm function.
Intended for use in conjunction with other medical devices
The monitor can be used together with high-frequency electrotomes and
defibrillators.
2.4 Safe Operation Conditions
Methods of sterilization or
disinfection recommended by
the manufacturer
Sterilization: not applicable
Disinfection: Refer to Maintenance and
Cleaning Chapter
Electromagnetic interference No mobile telephone nearby
Electrosurgical interference
damage
No damage
Diathermy instruments
influence
Displayed values and prints may be disturbed or
erroneous during diathermy
Defibrillation shocks The monitor specifications fulfill the
requirements of IEC 60601-1, IEC 60601-2-27,
IEC 60601-2-49, IEC 60601-2-34
Auxiliary outputs
The monitor must fulfill the requirements of
standard IEC 60601-1-1