Patient Monitor User’s manual
10-2
Note:
If Temp to be measured beyond probe’s measuring range, over
measuring range alarm will be displayed on the screen. Check if probe
is on the corresponding patient body site, or change it to other site on
the patient.
If “Temp self-check error” is displayed on the screen, it is possibly
that something is wrong with the temperature module, the operator
should stop using the module and contact the manufacturer.
10.3 Monitoring Procedure
1. Selecting Temp probe
Select the correct type and size of probe for your patient.
2. Connecting Temp probe to monitor
Plug the Temp cable into the Temp connector on monitor.
3. Applying the probe to patient
Apply the probe to the patient correctly.
4. Selecting Label
Select an appropriate temperature label.
5. Checking the Alarm Setting
Check that the alarm settings are appropriate for this patient and this type of
Temp measurement.
Warning: Make sure you set alarm limits for the correct label. The
alarm limits you set are stored for that particular label only.
Changing the label may change the alarm limits.