Patient Monitor User’s manual
11-5
11.7 Setting NIBP
11.7.1 Setting Unit
Select NIBP parameter area, and select【Unit】in the NIBP setting menu, and
The options are【mmHg】or【kPa】.
11.7.2 Selecting Alarm Source
Select the alarm source for NIBP parameters. When the selected parameter
or one of the parameter exceed alarm limit, the monitor will give out alarm
signal. The options are as follows:
【Sys】
:Only Systolic pressure exceeds the alarm limit will trigger the alarm
system.
【Dia】
:Only Diastolic pressure exceeds the alarm limit will trigger the alarm
system.
【Map】
:Only Mean pressure exceeds the alarm limit will trigger the alarm
system.
【Sys&Map】
:Systolic pressure or mean pressure exceeds the alarm limit will
trigger the alarm system.
【Map&Dia】
:Mean pressure or diastolic pressure exceeds the alarm limit will
trigger the alarm system.
【Sys&Dia】
:Systolic pressure or diastolic pressure exceeds the alarm limit will
trigger the alarm system.
【 Sys&Dia&Map 】
:Mean pressure, systolic pressure or diastolic pressure
exceeds the alarm limit will trigger the alarm system.
11.7.3 Setting Initial Cuff Inflation Pressure
Select NIBP parameter area, and select【Inflation】in the NIBP setting
menu, set the initial cuff inflation pressure according to patient type and
requirement.
11.7.4 Setting Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure, and block the
venous blood vessel to assist venous puncture.
1. Select the NIBP parameter area, and enter the NIBP parameter setting