Patient Monitor Users manual
11-6
menu.
2. SelectVenipunc. Press, and set to a proper value.
3. SelectStat/Stop Assist Venipunc.to start it.
4. Puncture vein and draw blood sample.
5. SelectStat/Stop Assist Venipunc.again to deflate the cuff. The cuff
deflates automatically after a set time if you do not deflate it.
11.7.5 NIBP Resetting
Select NIBP parameter area, and selectReset, then the inflation value of
blood pressure pump restores to the initial value. In case the blood pressure
pump doesnt work as normal but without any prompt, the blood pressure
pump can be checked by reset, thus the blood pressure pump in abnormal
condition due to unexpected reason will automatically restore.
11.8 NIBP Maintenance
The maintenance and calibration of NIBP measurement is necessary for
every two years. If you need to maintain NIBP, please contact the professional
service personnel. More details about NIBP maintenance, please refer to
Service Manual
.
.