Patient Monitor Users manual
5-4
Technical alarm
1) Technical alarm messages are displayed in the technical alarm area.
2) The * symbol before the alarm message match the alarm level as follows:
High level alarms: ***
Medium level alarms: **
Low level alarms: *
3) The system uses different background colors for the alarm message to
match the alarm level:
High level alarms: red
Medium level alarms: yellow
Low level alarms: blue
Prompt messages
1) Prompt messages are displayed in technical alarm area or the corresponding
parameter area.
2) Prompt messages have no color and visual and audible alarm indication.
When multiple alarms occur at the same time, the alarm messages will be
displayed in the alarm area in turn.
Select the physiological alarm or technical area, an alarm message
window will pop up, you can observe the alarm message from it.
5.3.4 Flashing Numeric
When a physiological alarm occurs, the numeric of parameter flashes.
5.4 Alarm Status Symbol
The system sound is silenced.
The alarm sound is off.
The alarm is suspended.
The parameter alarm is off.