Patient Monitor Users manual
5-6
5.6.3 Alarm Limit
Setting Individual Alarm Limits
The alarm limit can be set in each parameter setting menu, you can select
Hi Lmt, Lo Lmtto set the alarm limit for each parameter. The alarm
limit is adjustable, Physiological alarm will be triggered when the measuring
value exceeds the set limit.
Setting All Alarm Limits
SelectMain MenusmartkeyAlm SetupAlm Limitto set
the alarm limits of all present parameters.
Warning: Medical personnel should set the alarm limits of
parameters in line with the clinical environment and
existing clinical experience. Before monitoring, please
confirm whether the alarm setting is suitable for the
monitored patient.
5.7 Silence
SelectSilencesmartkey, all the sound of system is shut off. Select
Silencesmartkey once again to exit the silence status.
When in the silence status, the alarm indicators are valid except audible
alarm. The system sound includes alarm sound and QRS sound.
Note: When in the silence status, the icon will be displayed in the
upper right corner of screen.
5.8 Pausing Alarms
Press the button on the front panel of monitor, you can suspend all alarm
indicators of the monitor:
The visual alarm and audible alarm are all suspended.
The parameters of physiological alarm stop flashing.
The alarm message in the physiological alarm area will not be displayed.
The remaining time and the icon will be shown in the physiological
alarm area.