Patient monitor User’s manual
19-4
otherwise the recording meter may be damaged.
Unless for paper replacement or fault remedy, don’t keep the recorder
door open.
19.6 Clearing Jam Paper
While the sound of recorder operation or printing of recording meter is
abnormal, please first check whether there is paper jam in the recording meter.
If so, please clear it as per following steps:
1. Open the recorder door;
2. Pull out the recording paper, and cut off the wrinkle part;
3. Load recording paper once again and close the recording meter door.
19.7 Cleaning Recorder
After long-time service, some paper scrap and impurity will accumulate on
the printing head, and affect printing quality as well as the service life of
printing head and roll shaft. The recorder can be cleaned according to the
following methods:
1. Before cleaning, the measures such as wearing anti-static wrist strap
shall be adopted to avoid the damage to recording meter resulting
from static;
2. Open the recorder door and pull out recording paper;
3. Use a tampon with some alcohol to sweep slightly the surface of
thermo-sensitive parts of printing head;
4. After the alcohol entirely vaporizes, load recording paper once again
and close the recorder’s door.
Caution:
Don’t use any article that can damage the thermo-sensitive parts of
recorder during cleaning.
Don’t heavily press the printing head of recorder.