EasyManua.ls Logo

Biolight Q7 - Removing Exhaust Gases from the System; Safety Information

Default Icon
210 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor Users manual
13-9
measurements or other error conditions. If you attempt zeroing while
CO
2
remains in the adapter or cannula, the time required to zero the
sensor may be increased.
13.6 Removing Exhaust Gases from the System
Warning: When using the microstream CO
2
measurement on
patients who are receiving or have recently received
anesthetics, connect the outlet to a scavenging system,
or to the anesthesia machine/ventilator, to avoid
exposing medical staff to anesthetics.
Use an exhaust tube to remove the sample gas to a scavenging system.
Attach it to the outlet connector of microstream CO
2
sensor.
13.7 Safety Information
Warning:
Do not position the sensor cables or tubing in any manner that may
cause entanglement or strangulation.
Reuse, disassembly, cleaning, disinfecting or sterilizing the single
patient use cannula kits and on-airway adapters may compromise
functionality and system performance leading to a user or patient
hazard. Performance is not guaranteed if an item labeled as single
patient use is reused.
Inspect the microstream on-airway adapters, microstream sampling kits
and CO
2
airway adapters for damage prior to use. Do not use the
microstream on-airway adapters, microstream sampling kits and CO
2
airway adapters if they appear to be damaged or broken.
Replace the microstream on-airway adapters, microstream sampling
kits and CO
2
airway adapters if excessive secretions are observed.
Monitor the CO
2
waveform (Capnogram). If you see changes or
abnormal appearance check the airway adapters and the sampling tube.
Replace it if needed.
Monitor the CO
2
waveform (Capnogram) for elevated baseline. Elevated

Table of Contents

Related product manuals