Patient Monitor Users manual
10-1
Chapter 10 Temperature (Temp)
10.1 Introduction
The monitor measures Temperature with Temperature sensors.
The monitor has maximum eight channels for Temp measurement option,
and can display the temperature of eight channels and temperature difference
(TD) at the same time.
10.2 Safety Information
Warning:
Disposable Temp probes must not be re-sterilized or reused.
The calibration of temperature measurement is necessary for every
two years (or as frequently as dictated by your Hospital Procedures
Policy). When you need calibrate the temperature, please contact the
manufacturer.
Should the Temp probe become damaged beyond repair, or for some
reason its useful life is considered to be at an end, please observe all
local, state, and federal regulations that relate to the disposal of this
products or similar products.
Caution:
The self-test of the temperature measurement is performed
automatically termly during the monitoring. The test procedure lasts
about 1s and does not affect the normal measurement of the Temp
monitoring.
The Temp sensors and cables should be handled with care. When not
in use, the sensor and the cable should be rounded into loose ring
shape.