Patient Monitor User’s manual
3-10
you plug in another Temp module (module B) with the label Tskin, thus the
monitor will close the module B automatically.
To resolve conflict of labels:
—— You can keep on using module A to measure Tskin;
—— If you decide to measure Tskin with module B, you need to modify the
label Tskin with the module A first, and activate the module B once
again.
Other modules
The monitor can simultaneously support three independent IBP modules
and IBP module in EMS, also support three independent Temp modules and
the Temp module in EMS, but only support one measurement module of other
type, otherwise the monitor will close the second module automatically.
Take the CO
2
module as an example: a monitor has been equipped with
one CO
2
module (module A), furthermore, the module A is active, and at that
time, if you plug in another CO
2
module (module B), the monitor will close
the module B automatically.
To resolve the conflict:
——You can keep on using module A;
——If you need to use the module B, you must pull out both module A and B,
then plug in the module B. Or you can close the module A and activate
the module B in the【Module Set】window.
3.11 Using Mouse
The monitor supports USB-interface mouse and allows hot plugging and
plug & play.
While using mouse, you should take care of the following:
The left key of a mouse is the major key, and the right key of the mouse is
the secondary key.
To click the major key of a mouse is equivalent to the operation of
pressing trim knob.
3.12 Using Soft Keyboard
The software of the monitor can provide soft keyboard to input data and
support entering in Chinese and English.