Patient Monitor User’s manual
2-4
Attention! Please refer to the documents accompanying this
monitor (this manual)!
4. When a defibrillator is applied on a patient, the monitor may have transient
disorders in the display of waveforms. If the electrodes are used and placed
properly, the display of the monitor will be restored within 10s. During
defibrillation, please note to remove the electrode of chest lead and move the
electrode of limb lead to the side of the limb. The electrode of the defibrillator
should not come into direct contact with the monitoring electrodes. Please
ensure the monitor is reliably grounded and the electrodes used repeatedly
should be kept clean.
Warning: When conducting defibrillation, do not come into contact
with the patient, the bed and the monitor. Otherwise erious
injury or death could be resulted in.
5. To guarantee the safe operation of the monitor, the monitor is provided with
various replaceable parts, accessories and consuming materials (such as
sensors and their cables, electrode pads). Please use the products provided or
designated by the manufacturer.
6. Monitor only guarantees its safety and accuracy under the condition that it
is connected to the devices provided or designated by manufacturer. If the
monitor is connected to other undesignated electrical equipment or devices,
safety hazards may occur for causes such as the cumulating of the leakage
current.
7. To guarantee the normal and safe operation of the monitor, a preventive
check and maintenance should be conducted for the monitor and its parts
every 6-12 months (including performance check and safety check) to verify
the instrument can work in a safe and proper condition and it is safe to the
medical personnel and the patient and has met the accuracy required by
clinical use.
Caution: The monitor does not contain any parts for self-repair by
users. The repair of the instrument must be conducted by
the technical personnel authorized by manufacturer.