Patient Monitor User’s manual
7-9
【Diagnosis】: Use when diagnostic quality is required. The unfiltered ECG
wave is displayed so that changes of waveform are visible.
【Surgery】:Use while the signal is distorted by high-frequency or
low-frequency interference. During a surgery, selecting the mode【Surgery】
can reduce artifacts and the interference from electro-surgical units. Under
normal measurement conditions, selecting【Surgery】may suppress the QRS
complexes too much so as to interfere the ECG analysis.
【User】: Users can manually set the switches of various filters as required.
The states of the filter under various ECG modes
Filter
ECG mode
Drift filter HUM filter EMG filter
Diagnosis
Off Optional
Off
Surgery
Intense On Intense
Monitor
Mild On Mild
User
Optional
Optional
Optional
Note:Under the mode of【Surgery】and【Monitor】, the state of the filter
cannot be regulated. Only under the state 【Diagnosis】of【User】
can the state be regulated. Please select 【 Monitor 】 during
monitoring a patient, select【Surgery】under the state of great
interference.
Setting QRS Volume
The monitor will give out QRS sound according to HR value. Select【QRS
Volume】in the ECG setting menu to change the QRS volume. The volume can
be set to a value between 1 and 6, and also can be silenced.
Setting the Primary Lead
Select【Primary Lead】in ECG parameter setting menu, and you can set
the primary lead as required, the wave of primary lead will be displayed on
the top of the screen.
7.5.2 Setting ECG Waveform
Select one ECG waveform to enter the ECG wave setting menu.