ISE MODULE (optional)
Sample volume
Serum: 100 mL
Urine: 200 mL
ENVIRONMENTAL REQUIREMENTS
Room temperature 10° C to 35° C
10° C to 30° C (with ISE module)
Relative humidity < 85% with no condensation
Maximum height < 2 500 m
Contamination grade 2
Transportation and storage temperature 0° C to 40° C
Transportation and storage humidity < 85% with no condensation
DIMENSIONS AND WEIGHT
Dimensions (Width, depth and height) 1200 mm x 720 mm x 1258 mm
Weight 210 kg
ELECTRICITY REQUIREMENTS
Mains voltage 115 V to 230 V
Network frequency 50Hz or 60Hz
Electric power 500 VA
FLUID REQUIREMENTS
Water inlet rough external tank or through direct
connection
Type of water Puried type II (NCCLS)
Water consumption < 14 L/h
High contamination waste tank Internal, 5 L
Washing solution tank Internal, 5 L
MINIMUM COMPUTER REQUIREMENTS
Operating system Windows® 7 64 bit (x64)
CPU Equivalent to Intel Core i3@3.10 GHz or higher
RAM memory 4Gbytes
Hard disk 40Gbytes or more
DVD player Yes
VGA monitor Minimum resolution 1 024 x 768
Serial channel connector USB
COMPLIANCE WITH DIRECTIVES
EC-IVD Directive 98/79/EC