Recommended Maintenance
Schedule | 119
Recommended Maintenance Schedule
Keypad Users: The protocol names onboard the washer do not include the W- prefix and
some protocols include the model name, e.g. 405HT or 405Select, when appropriate.
Tasks
Frequency
Daily Overnight/
Multi-Day
Weekly Periodic/
Monthly
Before
storage/
shipment
Run W-DAY_RINSE
ü ü
Run AutoPrime
ü
Run W-OVERNIGHT_LOOP
ü
Run W-RINSE_AND_SOAK
ü
Run Verify™ clog detection test
ü
Run AutoClean
ü ü
Components
Remove protein residuals and
fungi growth, (if necessary)
ü ü ü
Check/empty waste bottles
ü ü
Clean bottles
ü ü
Clean plate carrier system
ü ü
Clean washer manifold
ü ü
Clean aspirate and dispense tubes
ü ü
Clean exterior surfaces and mist
shield
ü
Clean fluid inlet filter
ü ü
Clean vacuum filtration system
ü ü
Clean the Verify™ level sensor
ü ü
Decontaminate
Decontaminate external surfaces
ü
ü
Run W-DECONTAMINATE
ü ü
Decontaminate vacuum filtration
system
ü