200 | Chapter A: Error Codes
405 LS-LHC Software Error Codes
Generally, these errors are caused by protocol parameters that conflict with the
instrument's onboard settings. The protocol may have been originally created for a
different hardware configuration, the 192-tube wash manifold instead of the 96-
tube, for example.
Quick Fix: Make sure your Instrument Settings accurately reflect your
instrument’s hardware configuration and then, modify the protocol to fix any
invalid parameters. With the 405 LS connected to and communicating with your
computer and its main menu displayed on the keypad:
1. Click the Settings link in the main view.
2. In the Instrument Settings dialog, click the instrument link to get the settings
from the instrument.
3. Modify the protocol step that generated the error message.
Error
Code
Description Help
6000 General communication error during
download.
See Communication Errors on page 188
6001 COM port created by USB converter
no longer active
See Communication Port
6002 Invalid basecode part number;
instrument is not an 405 LS
Service Only.
Contact BioTek TAC.
6003 Invalid Basecode Data Version;
basecode needs to be updated
Contact BioTek to obtain latest basecode.
6004 No rows are selected for the specified
plate type
Modify the protocol to select a row.
6005 Invalid row selection value (must be
0 or 1)
Contact BioTek TAC.
6006 This instrument can only process 96-
well plates
The protocol may have been created for
another instrument, change the plate type
or select another protocol.
6007 This instrument can only process
1536-well plates
6008 The 8-tube Syringe Manifold can only
be used with 96-well plates
Mismatch between installed hardware and
protocol parameters: change the plate
type or correct the instrument settings to
match the currently installed hardware.
405 Microplate Washer LS