192 | Chapter A: Error Codes
System Error Codes
Most of these error conditions require technical expertise to correct. Error code 306
and few other exceptions to this rule are listed with remedies in the
Troubleshooting section. A few other errors may be caused by an obvious
obstruction to a device's movement or insufficient fluid in a supply vessel. Fix
these kinds of errors and restart your instrument to give it an opportunity to clear
the error code.
Contact BioTek Technical Assistance Center (TAC) for assistance.
Code Message What to do
100 Task was aborted Restart instrument if this message is
unexpected.
210, 220 Carrier X motor didnt find
home opto sensor transition
Carrier X motor didnt find
autocal jig opto sensor
transition
Clean the plate carrier, rails, and glide
strips, using mild detergent and hot water,
70% isopropyl alcohol or ethanol. Restart
the instrument. If the error occurs again,
contact BioTek TAC.
211, 221 Carrier Y motor didnt find home
opto sensor transition
Carrier Y motor didnt find
autocal jig opto sensor
transition
Run self test. If error reoccurs, contact
BioTek TAC.
212, 222 Dispense head motor didnt find
home opto sensor transition,
Dispense head motor didnt find
autocal jig opto sensor
transition
Run self test. If error reoccurs, contact
BioTek TAC.
213, 223 Wash head motor didnt find
home opto sensor transition,
Wash head motor didnt find
autocal jig opto sensor
transition
Run self test. If error reoccurs, contact
BioTek TAC.
220 Carrier X motor didnt find
autocal jig optical sensor
transition.
Service Only.
Contact BioTek TAC.
221 Carrier Y motor didnt find
autocal jig optical sensor
transition.
Service Only.
Contact BioTek TAC.