EasyManua.ls Logo

Biotek 405 LS - Storing the Instrument; Replace Components

Biotek 405 LS
240 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
144 | Chapter 4: Maintenance
With Buffer
Switching
Connect the supply bottles this way:
Buffer bottles C and D: Disinfectant
Buffer bottles A and B: Rinse solution
After disinfecting channels C and D, you will reverse this assignment
to disinfect channels A and B.
Keypad:Run DECON_STEP1, DECON_STEP2, and DECON_
STEP3.
LHC: At the prompt/delay, change bottles.
Without Buffer
Switching
Connect the disinfectant bottle to the Fluid In port.
Keypad:Run DECON_STEP1 followed by DECON_STEP2.
LHC: At the prompt/delay, change bottles.
7. Finally, disconnect the supply tubing from the washer to completely purge the
system of all fluid by priming the tubing with air. Keypad: Run PURGE_
SYSTEM_WITH_AIR. LHC: Follow the prompts.
Storing the Instrument
After performing the Long Shutdown (Prepare for Storage or Shipment) protocols:
l Turn off the instrument and disconnect the power cord.
l Store it on a flat surface that is relatively free of vibration, in a dust-free and particle-
free environment.
l Protect the instrument from temperature extremes that can cause condensation within
the unit and from corrosive fumes and vapors.
l Store the instrument under the following environmental conditions:
Temperature: 2 to 50°C (-4° to 122°F)
Relative humidity: 10% to 85% (non-condensing)
n Important: Allow the instrument to reach room temperature before use after
storage.
Replace Components
Some components of the 405 LS must be replaced periodically to maintain specified
performance levels.
l Replace the Vacuum Pump Fuse on the facing page
l Clean the Fluid Inlet Filter on page 130
l Vacuum Filtration Carrier: Replace the Gasket on page 146
405 Microplate Washer LS

Table of Contents

Related product manuals