AutoPrime | 123
LHC: Keypad:
4. Click Send to transfer the
settings to the instrument.
5. Define the AutoPrime
parameters: rate, volume, and
buffer valve, if applicable. Press
Enter at each screen to advance
to the next.
AutoPrime for Overnight-Weekend Maintenance
AutoPrime can be used to keep the manifold tubes wetted during idle periods
throughout the day and then modified to soak the tubing for longer periods
overnight and on the weekends. See AutoPrime on previous page for setup
instructions.
AutoPrime Parameters Washer
AutoPrime Interval 15 minutes
Device Washer
AutoPrime Yes/Enabled
Volume 90 mL
Rate 09
Buffer Any
Submerge Duration 120 minutes/2 hours
Turn on AutoPrime for the Washer
l Learn about AutoPrime, if you haven't already done so.
n Keypad users: Select UTIL at the main menu, then AUTPRM and specify
parameters.
1. Select Tools>Instrument Utilities>AutoPrime.
2. Fill the Enabled checkbox.
3. Set the AutoPrime Interval: enter the downtime interval that will trigger an
AutoPrime.
4. Select the Buffer to use if you have Buffer Switching installed.
5. Set the Volume and Rate. The default values are 40 mL at flow rate 9.