System Error Codes | 193
Code Message What to do
223 Wash head motor didn’t find
autocal jig optical sensor
transition.
Service Only.
Contact BioTek TAC.
300 Carrier X motor interlock safety
switch open
Service Only.
Contact BioTek TAC.
301 Carrier Y motor interlock safety
switch open
Service Only.
Contact BioTek TAC.
302 Dispense head motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
303 Wash head motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
400 Carrier X motor failed positional
verify
Run self test. If error reoccurs, contact
BioTek TAC.
401 Carrier Y motor failed positional
verify
Run self test. If error reoccurs, contact
BioTek TAC.
402 Dispense head motor failed
positional verify
Run self test. If error reoccurs, contact
BioTek TAC.
403 Wash head motor failed
positional verify
Service Only.
Contact BioTek TAC.
600-606 Specified motor currently in use Service Only.
Contact BioTek TAC.
700 -
70F
Invalid motor number specified Service Only.
Contact BioTek TAC.
900 Calibrationon failed. The
measured or calculated autocal
value is out of tolerance
Service Only.
Contact BioTek TAC.
A00 Invalid plate type selected The currently selected plate type is not
supported with the currently installed or
requested hardware. If this is not the case,
contact BioTek TAC.
C01 Configuration or autocal data
missing
Service Only.
Contact BioTek TAC.
C02 Checksum mismatch -calculated
checksum didn’t match saved
checksum
Service Only.
Contact BioTek TAC.