194 | Chapter A: Error Codes
Code Message What to do
C03 Configuration parameter out of
range
Service Only.
Contact BioTek TAC.
1001 Bootcode powerup checksum
test failed
Contact BioTek TAC.
1002 Unknown error in bootcode Contact BioTek TAC.
1003 Bootcode page program error Contact BioTek TAC.
1004 Bootcode block size error (not
256)
Contact BioTek TAC.
1005 Invalid processor signature (not
1280,1281,2560,2561)
Contact BioTek TAC.
1006 Bootcode memory exceeded Contact BioTek TAC.
1007 Invalid slave port Contact BioTek TAC.
1008 Invalid response from slave Contact BioTek TAC.
1009 Invalid processor detected Contact BioTek TAC.
1010 Checksum error downloading
basecode
Contact BioTek TAC.
1250 Internal RAM test error on the
UI processor
Contact BioTek TAC.
1251 Internal RAM test error on the
MC processor
Contact BioTek TAC.
1260 Stack test error on the UI Contact BioTek TAC.
1261 Stack test error on the MC Contact BioTek TAC.
1400 No vacuum pressure detected
after turning on the vacuum
pump
Make sure the vacuum pump is turned on,
and not leaking; the waste bottle has not
overflowed and its caps are seated
correctly. If the pump is not running, the
fuse may be blown. But, before replacing the
fuse, first try to determine and remedy the
cause of the failure. If the pump sounds
strained or runs slowly, it may be damaged.
Review other related Troubleshooting
suggestions.