System Error Codes | 195
Code Message What to do
1401 The waste bottles should be
emptied before continuing
Empty the waste bottles and rerun the
protocol.
1402 The requested valve is invalid Contact BioTek TAC.
1403 Invalid Z-axis offset value The Magnet Adapter offset (MAGHT) is too
high, the "in use" value is missing, or a more
serious software problem exists. Make sure
the setting is valid, and if so, contact BioTek
TAC.
1404 Plate type restricted This instrument model or the requested
device does not support the selected plate
type. Edit the protocol to change the plate
type.
1405 Z-axis out of range The requested travel/dispense height cannot
be reached. The conflict may be caused by a
combination of variables, plate type, Plate
Clear or other height settings. Review the
protocol parameters and instrument settings
to identify a correction.
1407 Invalid step type The protocol may have been created for a
different instrument, it is not compatible
with this instrument.
1408 Invalid plate geometry Contact BioTek TAC.
1409 Invalid plate carrier type Contact BioTek TAC.
1410 Incompatible hardware Vacuum filtration requested but either the
instrument is in BioStack mode, or this
instrument is not configured to support it.
Change the BioStack configuration, it is not
compatible with the vacuum carrier, or
contact BioTek TAC.
1411 Invalid plate carrier The vacuum filtration carrier must be
installed to run this kind of protocol, and the
plate carrier setting must represent the
currently installed carrier.
1412 Tip clearance error Contact BioTek TAC.
1413 AutoPrime in progress Contact BioTek TAC.
1414 AutoPrime aborted Contact BioTek TAC.