Dispense Precision Test Worksheet
96-Tube Manifold
Annual Buffer Switching Test
Serial Number:
Calculations for Valve A
Standard Deviation:
Mean OD (Sum of all wells ÷ Number of wells):
% CV (Standard Deviation ÷ Mean OD x 100):
% CV <= 3.0? q Pass q Fail
Calculations for Valves A-D
(Annual OQ only, check here q if not performed)
Total Dispense Weight 28.8 g ±10%?
(25.92 g - 31.68 g)
Valve A grams q Pass q Fail
Valve B grams q Pass q Fail
Valve C grams q Pass q Fail
Valve D grams q Pass q Fail
Date:
Test Performed By:
405 LS Qualification Procedures