System Error Codes | 197
Code Message What to do
1513 Plate type - manifold conflict Change the Plate Type to one supported by
the manifold. Or, click the Instrument
Settings link and make sure they match the
physical hardware: Get settings from
instrument.
1514 Ultrasound module not
connected
The protocol may have been created for a
different instrument, according to the
instrument's basecode it is not compatible
with this instrument. Contact BioTek TAC.
1515 Cell Wash hardware not
installed
1516 Vacuum filtration start or end
error
Vacuum pressure is detected in the
intermediate waste bottle before or after the
run. Conatct BioTek TAC.
1517 Vacuum filtration sensor error Unable to read vac. filtration sensor. This is
expected if you are running with very low
pressure, in this case disable the sensor.
Otherwise, check for leaks, e.g. bottle cap.
1600-
160C
The onboard storage space
allocated for this function has
been used up.
Use the LHC ”Manage Memory” control to
reallocate space.
160D Not a valid step Contact BioTek TAC.
1700-
1709
"Q" model level sensor errors Contact BioTek TAC.
170A-
170F
"Q" model level sensor RX
errors
Contact BioTek TAC.
1710-
1716
"Q" model level sensor RX
errors
Contact BioTek TAC.
1717-18 Level sensor X-axis parameter
out-of-range
Contact BioTek TAC.
171A-
171F
Level sensor initiation errors Contact BioTek TAC.
1720-
1736
Level sensor functional errors Take action to correct the error, if possible.
Otherwise, contact BioTek TAC for
assistance.
1737 Plate test did not find plate. Put a plate on the plate carrier.