20 | Chapter 2: Installation
Waste System
Three lengths of tubing are shipped with the waste module:
Tubing: Connects:
Short tube with yellow and green bands
g
The two waste bottles to each other
Long tube with green bands on both ends
g
Bottle without sensor to vacuum port
Long tube with yellow and orange bands
g
Bottle with waste sensor to the vacuum
pump
1. Locate the quick-release caps shipped inside the waste bottles and attach the
tubing to them as follows:
2. Connect the waste bottles to each other using the shortest length of tubing,
matching the colored bands on the tubing to colored dots on the caps.
3. Attach the waste sensor cable to the Waste Sensor port on the back of the
washer.
4. Attach the tube from the waste bottle with the waste sensor in its cap to the
vacuum pump.
405™ Microplate Washer LS