Verify Performance | 31
Run the Verify test:
LHC
1. Select Tools>Instrument Utilities>Verify Manifold and click Run a new Test.
l Buffer: Select the bottle to use, if applicable.
l Prime: optionally, prime the manifold tubes to correct for evaporation loss.
40 mL of fluid is dispensed.
l %CV Threshold: optionally, change the %CV to match your lab's standard for
QC tests: 5-15%. The default value of 5% is recommended.
2. Press Start to run the test.
3. Assess the results: Verify Results.
Do not discard the Verify Test Plate. Reuse it for as long as possible. See
Replacement Procedure for Verify Test Plate on page 134 if it is damaged.
Contact BioTek TAC if any of the tests fail.