Aesculap
®
ELAN 4 air motor hose 3 m with hand control GA706R or ELAN 4 air motor
hose 5 m with hand control GA707R, ELAN 4 air wall adapters GA710R to
GA715R
Validated cleaning and disinfection procedure
7.6 Manual cleaning and wipe disinfection
D–W: Drinking water
FD–W: Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
RT: Room temperature
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. To avoid the risk of corrosion/malfunctioning,
allow any fluid that has entered the product to drain out immediately.
Phase I
Clean the product under running tap water, using a suitable cleaning
brush until all visible residues have been removed from the surfaces.
Mobilize non-rigid components, such as set screws and hinges, during
cleaning.
Phase II
Dry the product in the drying phase with suitable equipment (e.g. cloth,
compressed air).
Phase III
Wipe all surfaces of the product with a single-use disinfecting wipe.
Phase IV
After the specified exposure time (at least 1 min), rinse the disinfected
surfaces under running FD water.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable equipment (e.g. cloth,
compressed air).
After manual cleaning/disinfection, check visible surfaces visually for
residues.
Repeat the cleaning/disinfection process if necessary.
Product Validated procedure Reference
Motor hose with
hand control
Manual cleaning and wipe
disinfection
see Manual clean-
ing and wipe disin-
fection
Manual pre-cleaning with
brush and subsequent
mechanical alkaline cleaning
and thermal disinfection
see Mechanical
cleaning/disinfec-
tion with manual
pre-cleaning
Wall adapter Wipe disinfection without
sterilization
see Wipe disinfec-
tion
Phase Step T
[°C/°F]
t
[min]
Conc.
[%]
Water quality Chemicals
I Cleaning RT (cold) - - D–W -
II Drying RT - - - -
III Wipe disinfection - >1 - - Meliseptol HBV wipes 50 % Propan-1-ol
IV Final rinse RT (cold) 0.5 - FD–W -
VDrying RT - - - -